Work

A linear mixed perfusion model for tissue partial volume correction of perfusion estimates in dynamic susceptibility contrast MRI: Impact on absolute quantification, repeatability, and agreement with pseudo-continuous arterial spin labeling

Year: 2016
Type: article
Abstract: PURPOSE: The partial volume effect (PVE) is an important source of bias in brain perfusion measurements. The impact of tissue PVEs in perfusion measurements with dynamic susceptibility contrast MRI (DSC-MRI) has not yet been well established. The purpose of this study was to suggest a partial volume correction (PVC) approach for DSC-MRI and to study how PVC affects DSC-MRI perfusion results. METHODS: A linear mixed perfusion model for DSC-MRI was derived and evaluated by way of simulations. Twenty healthy volunteers were scanned twice, including DSC-MRI, arterial spin labeling (ASL), and partial volume measurements. Two different algorithms for PVC were employed and assessed. RESULTS: Simulations showed that the derived model had a tendency to overestimate perfusion values in voxels with high fractions of cerebrospinal fluid. PVC reduced the tissue volume dependence of DSC-MRI perfusion values from 44.4% to 4.2% in gray matter and from 55.3% to 14.2% in white matter. One PVC method significantly improved the voxel-wise repeatability, but PVC did not improve the spatial agreement between DSC-MRI and ASL perfusion maps. CONCLUSION: Significant PVEs were found for DSC-MRI perfusion estimates, and PVC successfully reduced those effects. The findings suggest that PVC might be an important consideration for DSC-MRI applications. Magn Reson Med 77:2203-2214, 2017. © 2016 International Society for Magnetic Resonance in Medicine.
Institution: Lund University
Language: English
Field: Medicine
Open Access status: bronze
OpenAlex ID: W2467223444
FWCI: 0.503
Cites: 31
Cited by: 8
Related to: 10