Work

The Compton source efficiency tracing method, a new standardization method in liquid scintillation counting

Year: 2008
Type: article
Abstract: We describe a new standardization method in liquid scintillationcounting based on the use of a temporary virtual tracer source createdinside the scintillator by Compton interaction. The Compton tracersource is measured by the triple coincidence liquid scintillation counterafter selection of the Compton events by a gamma-ray detector. Thepaper describes the principle of the method, the experimental setup andpresents the results obtained by this method for the standardization of a3H solution. These later results are compared with those obtained usingthe classical Triple to Double Coincidence Ratio method. Possibleapplication of this new method for international comparisons isaddressed.
Source: HAL
Language: English
Subfield: Radiation
Open Access status: green
OpenAlex ID: W2482019242
FWCI:
Cites:
Cited by:
Related to: 10