Design, construction and testing of a low-cost automated (68)Gallium-labeling synthesis unit for clinical use.
Year: 2016
Type: article
Abstract: The interest in (68)Gallium labeled PET probes continues to increase around the world. Widespread use in Europe and Asia has led to great interest for use at numerous sites in the US. One barrier to entry is the cost of the automated synthesis units for relatively simple labeling procedures. We describe the construction and testing of a relatively low-cost automated (68)Ga-labeling unit for human-use. We provide a guide for construction, including part lists and synthesis timelists to facilitate local implementation. Such inexpensive systems could help increase use around the globe and in the US in particular by removing one of the barriers to greater widespread availability. The developed automated synthesis unit reproducibly synthesized (68)Ga-DOTATOC with average yield of 71 ± 8% and a radiochemical purity ≥ 95% in a synthesis time of 25 ± 1 minutes. Automated product yields are comparable to that of manual synthesis. We demonstrate in-house construction and use of a low-cost automated synthesis unit for labeling of DOTATOC and similar peptides with (68)Gallium.
Source: PubMed
Authors: Pedram Heidari, Alicia Szretter, Laura E Rushford, Maria Stevens, Lee Collier, Judit Sore, Jacob Hooker, Umar Mahmood
Institution: Massachusetts General Hospital
Language: English
Subfield: Radiology, Nuclear Medicine and Imaging
Field: Medicine
Domain: Health Sciences
Open Access status: green
Funder: National Institutes of Health
OpenAlex ID: W2509958868
