Abstract: Atmospheric chemistry investigations often require a multitude of measurements which can be obtained only through the utilization of airborne sampling platforms. Instrument limitations and the aircraft environment present several considerations for sampling-system design, including such factors as instrument sensitivities and response times, altitude effects, sampling intervals for acquiring samples, and physical compatibility with the aircraft. An aircraft system with an extensive evolutionary instrument array has been in development at PNL for several years during which several special systems have been developed to improve aircraft measurement capabilities. A high-volume air sampling system providing flows of up to 4 m/sup 3//min and simultaneous collection of three filters in parallel has been constructued to reduce filter collection times. A constant pressure inlet system was developed to overcome adverse effects in instrument response resulting from altitude changes. The system functions so that instruments which are connected experinece a constant pre-set pressure regardless of the sampling altitude. This system is particularly useful for airborne operation of a flame photometric sulfur analyzer. Special chemiluminescence NO/NO/sub x/ analyzers utilizing photon counting were built which are capable of fast response and detecton of concentrations in the sub-ppB range.