Longitudinal measurements of total body water and body composition in healthy volunteers by online breath deuterium measurement and other near-subject methods.

Year: 2004
Type: article
Study design: Observational Study
Abstract: = 0.96-0.98), slope 1.02-1.03, CV = 4.5-11.6%. Systematic errors between the different methods in determining TBW were effectively identical at baseline and after one year. There was a tendency for subjects to gain weight during the study period, due to an increase, predominantly in younger women, of body water (FA-MS and SFT) and loss of upper body fat (ultrasound). BIA was relatively insensitive to these changes. It is concluded that over a 12-month period, TBW determined by FA-MS deuterium breath analysis has reproducibility similar to conventional weighing. The stability of between method errors would suggest that these techniques might be used in conjunction with each other in the longitudinal determination of body composition and so detect relatively subtle changes. The value of including an absolute determinant of TBW by FA-MS that is independent of the need to employ population derived equations, appears to be of value in the near-subject determination of body composition as required in clinical practice.