Factors limiting the linearity of response of tissue equivalent proportional counters used in micro- and nano-dosimetry
Year: 2017
Type: article
Abstract: Proportional counters filled with tissue equivalent gas mixtures are extremely useful instruments and are being used extensively as sensitive detectors for all types of radiations to measure the energy transferred to small tissue volumes. The linearity of their response is of primary importance. So the investigation and clarification of the physical phenomena taking place in the counter and of the limits within which useful results may be obtained would contribute to a more efficient use and a wider application of these counters. The linearity of response in the dose and in the gas gain has been determined. Linearity in the dose is limited by the total count rate effect, while linearity in the gas gain is limited by secondary processes occurring in the electron avalanche and by the self-induced space charge effect.