Work

Consistency of the national realization of dew-point temperature using NIS standard humidity generators

Year: 2017
Type: article
Abstract: A comparison of two standard humidity generators (two-temperature (2-T) and one-temperature (1-T) generators) that are developed by the National Institute for Standards (NIS) has been performed using a transfer standard chilled-mirror hygrometer and measurement procedures to realize dew-point temperature Td in the range from −50 °C to 0 °C. The main objective of this comparison was to compare the realizations of dew-point temperature and to establish the level of consistency between the two generators. For a level of consistency between two measurements, it is expressed by the difference between the measured values, m1 − m2, and the expanded pair uncertainty of this difference Up [1]. The comparison measurements revealed dew-point temperature differences of 0.02 °C and 0.07 °C with expanded pair uncertainties of ±0.09 °C and ±0.15 °C.
Language: English
Subfield: Instrumentation
Open Access status: gold
Estimated APC paid: $565
OpenAlex ID: W2582140832
FWCI: 4.526
Cites: 9
Cited by: 23