Work

Systolic blood pressure estimation using PPG during physical exercise

Year: 2016
Type: conference-paper
Abstract: A model to estimate systolic blood pressure (SBP) using finger-derived photoplethysmography (PPG) signals is proposed for 19 healthy subjects during physical exercise. The estimation was done for every 30 s during the period of exercise, before which an initialization process for each subject at rest was performed. Eighteen features were extracted from PPG signals and a multiple linear regression (MLR) method was employed to combine all derived features to estimate SBP. Our results showed that the proposed model had a performance in estimating SBP with a mean difference of -0.40 mmHg, a standard deviation of difference of 14.07 mmHg, and a median intra-subject correlation coefficients of 0.85. The present work demonstrated promising results of estimating SBP using PPG during physical exercise.
Language: English
Open Access status: green
OpenAlex ID: W2586468660
FWCI: 0.472
Cites:
Cited by: 5
Related to: 10