Work

Quantitative Assessment of Variation in CT Parameters on Texture Features: Pilot Study Using a Nonanatomic Phantom

Year: 2017
Type: article
Abstract: SUMMARY: Our aim was to evaluate changes in texture features based on variations in CT parameters on a phantom. Scans were performed with varying milliampere, kilovolt, section thickness, pitch, and acquisition mode. Forty-two texture features were extracted by using an in-house-developed Matlab program. Two-tailed t tests and false-detection analyses were performed with significant differences in texture features based on detector array configurations (Q values = 0.001–0.006), section thickness (Q values = 0.0002–0.001), and acquisition mode (Q values = 0.003–0.006). Variations in milliampere and kilovolt had no significant effect.
Language: English
Field: Medicine
Open Access status: hybrid
OpenAlex ID: W2603308445
FWCI: 5.243
Cites: 28
Cited by: 53
Related to: 10