Calibration of Glass Fiber Microcantilevers

Year: 2010
Type: article
Abstract: A thermal fluctuation method for microcantilever spring constant measurement was evaluated.Using this technique, the spring constant was measured by recording microcantilever movement under microscope, then analyzing a set of images, frame by frame to obtain the variance of the fiber's fluctuating position which can be related to the spring constant using the Equipartition Principle.Images were captured from VCR recordings as well as directly from the camera.Captured images were processed and fiber centroids were found using LabVIEW programs.The centroid data was processed using a moving average filter and variance was calculated for final determination of spring constant.Under some circumstances and when the fiber does not stick to the chamber wall, the thermal fluctuation method gives a spring constant that varies with cantilever length and is of the correct order of magnitude.Spring constant estimates using 40x magnification were always bigger than 20x for calculated spring constant values.Liquids have a much larger damping effect on vibrations compared to air for the spring constant measurement and are thus less affected by outside disturbance.Recording to VCR does not introduce significantly more noise from direct capture from the camera.
Author: Yang Liu
Institution: Ohio University
Language: English
Open Access status: green