MRD detection in multiple myeloma: comparison between MSKCC 10-color single-tube and EuroFlow 8-color 2-tube methods
Year: 2017
Type: article
Abstract: = 0.97). The overall concordance between the 2 tests was 98%. In 1 case, there was a discrepancy near the limit of detection of both tests in favor of the slightly greater theoretical sensitivity of the EuroFlow 8-color 2-tube panel (analytical sensitivity limit of MSKCC single 10-color tube: 6 cells in 1 million with at least 3 million cell acquisitions; EuroFlow 8-color 2-tube panel: 2 cells in 1 million with the recommended 10 million cell acquisitions).
Source: Blood Advances
Language: English
Subfield: Hematology
Field: Medicine
Domain: Health Sciences
Open Access status: gold
Estimated APC paid: $3,800
OpenAlex ID: W2610736495