Implementing vertical resolution and in-situ manipulation of samples within a scanning electron microscope : the Nanofinger®
Year: 2016
Type: dissertation
Abstract: This thesis is about the development of a tool to implement vertical resolution within a scanning electron microscope (SEM). This becomes necessary as due to the principle of an SEM it is only possible to get vertical resolution by proper focussing, which is limited to some micrometers. As it is important to get access to height information of a sample during several processes in the SEM, especially when performing nanomanipulation, an additional tool is inserted into the SEM to provide the missing information. This so called Nanofinger® was originally invented by Klocke Nanotechnik GmbH in Aachen. It is mounted on a three-axis manipulator and operates similar to an atomic force microscope (AFM) to perform linescans on a given structure or to simply retrieve surface coordinates at certain positions on the sample.As the former Nanofinger® did not fulfill the requirements, a new concept to improve the Nanofinger® was established. The new Nanofinger® was realized to operate in the SEM. To succeed in this project, first, the issues of the old Nanofinger® have been investigated to develop a concept for the improvements. In the following part of this thesis, the development and construction of the new Nanofinger® is described. This includes the new scanhead as well as the fabrication of the sensors. The most important aspect of the optimization is the achieved vertical resolution. The noise in der vertical direction is 1 nm, limited by the manipulator. Moreover, structures of 2 nm could be resolved. A slim geometry with a height of 3 mm allows the operation of the Nanofinger® at low working distances under the SEM. Also, the Nanofinger® can be operated easier than before, allowing users with little knowledge about the AFM to use the Nanofinger®. This includes a distinct resonance spectrum of the sensors and a simple tip exchange. Besides of the hardware, also the software was improved to add additional functionality to the Nanofinger®. This is a feedback loop which is directly designed for this special tool. Moreover, the electrostatic force microscopy (EFM) mode was implemented. To demonstrate the improved performance of the new Nanofinger® beyond the quality of the aquired linescans, local anodic oxidation (LAO) of graphene achieved with the Nanofinger® is shown in chapter 5. After a brief overview of LAO of graphene, the sample preparation of monolayer graphene on hexagonal boron nitride (hBN) is described. Attention was paid on investigating graphene within the SEM without contaminating the sample. The oxidation of graphene was successfully performed under vacuum conditions, achieving a linewidth down to 50 nm. The LAO was done within the SEM, supplying additional water up to a pressure of $3\times 10^{-5}$ mbar and a tip sample voltage of 10 V. Insulation of the cuts has been shown by directly contacting the graphene with the Nanofinger® and by doing EFM measurements within the SEM.
Source: RWTH Publications (RWTH Aachen)
Language: English
Subfield: Atomic and Molecular Physics, and Optics
Field: Physics and Astronomy
Domain: Physical Sciences
Keywords: scanning electron microscopy, vertical resolution, nanomanipulation, atomic force microscopy, electrostatic force microscopy, local anodic oxidation, graphene, hexagonal boron nitride, monolayer graphene, tip fabrication, resonance spectrum, feedback control, sample contamination, nanofabrication
Open Access status: green
OpenAlex ID: W2615296249