Work

Active cavity absolute radiometer based on high-Tcsuperconductors

Year: 1998
Type: article
Abstract: To implement the detector-based radiometric scale in the new Medium Background Infrared (MBIR) facility at the National Institute of Standards and Technology (NIST), we have developed an electrical-substitution cavity radiometer that can be operated just above liquid-nitrogen temperature. This MBIR active cavity radiometer (ACR) utilizes a temperature-controlled receiver cone and an independently temperature-controlled heat sink. Being a thermal-type detector, low noise and drift of the radiometer signal depends mainly on low-noise temperature control of the receiver and heat sink. Using high critical-temperature ( T c ) superconducting thin-film temperature sensors in the active control loops, we have achieved closed-loop temperature controllability of better than 10 µK at 89 K for a receiver having an open-loop thermal time constant of about 75 s. For a flux level of 1 µW to 10 µW, the rms noise floor over a measurement cycle time is below 20 nW. This is the lowest noise level yet reported for a liquid-nitrogen-cooled electrical-substitution radiometer, and it is the first demonstration of the use of high- T c superconductors in such a radiometer. Potential uses for this ACR in the MBIR facility include absolute measurement of the broadband radiance of large-area 300 K cryogenic black-body sources, and absolute measurement of the spectral radiance of laser-illuminated integrating spheres for improved spectral responsivity measurements of infrared transfer standard radiometers.
Source: Metrologia
Language: English
Open Access status: bronze
OpenAlex ID: W2085411166
FWCI: 4.651
Cites: 7
Cited by: 24
Related to: 8