The Influence of Source-to-Image Distance on Effective Dose and Image Quality for Mobile Chest X-ray Imaging
Year: 2017
Type: book-chapter
Abstract: Aim: \nTo investigate the effect of source-to-image distance (SID) on effective dose \nand image quality at a fixed mAs. Furthermore, to determine whether effective dose \nvaries as SID is varied after compensating mAs for the inverse square law.\nMethod: \nA chest phantom with a simulated pathology was imaged at varying SIDs \nat a fixed mAs and kV. Observers visually compared the experimental images with \na reference image using criteria divided into 3 categories: anatomy, noise and \nnodule image quality. A 2AFC program was used to display the images. A 5-point \nLikert scale was used ranging from ‘much worse’ to ‘much better’. PCXMC 2.0 \n84\nwas used to estimate effective dose. At each SID, a new mAs was calculated using \ninverse square law correction. Effective dose was estimated again.\nResults: \nModal response for 2AFC visual evaluation of anatomy image quality was \n‘equal’ for every SID, implying no observable change. Modal response for 2AFC \nvisual evaluation of noise changed from ‘better’ to ‘worse’ as SID increased. Modal \nresponse for visual evaluation of nodule image quality was ‘worse’ at the smallest \nSIDs and ‘better’ at the largest SIDs. Decreasing SID from 190-100 cm: increased \neffective dose by 247%; compensating mAs reduced the effective dose by 36%.\nConclusion: \nFor variable SIDs used in this study, there is little perceived difference \nin anatomical image quality. Nodule image quality may decrease at shorter SIDs \nwhile effective dose increases. Therefore, when shorter SIDs are necessary due to \nlack of space, practitioners must be aware of patient dose.
Authors: Hauge, Ingrid Helen Ryste, Beecher, Donal, Ellermann, Anne, Hughan, Sophi, Nguyen, Thi, Parrone, Luca, Selboe, Silje, Soares, Flavio augusto, Van Wijngaarden, Stephan, Sanderud, Audun
Language: English
Topic: Radiation Dose and Imaging
Subfield: Radiology, Nuclear Medicine and Imaging
Field: Medicine
Domain: Health Sciences
Open Access status: green
Funder: University of Salford Manchester
OpenAlex ID: W2622201722