Noise Suppression in Unshielded Magnetocardiography: Least-Mean Squared Algorithm versus Genetic Algorithm
Year: 2012
Type: article
Abstract: This paper discusses adaptive noise cancellation in magnetocardiographic systems within unshielded environment using two algorithms, namely, the Least-Mean Squared (LMS) algorithm and the Genetic Algorithm (GA). Simulation results show that the GA algorithm outperforms the LMS algorithm in extracting a weak heart signal from a much-stronger magnetic noise, with a signal-to-noise ratio (SNR) of -35.8 dB. The GA algorithm displays an improvement in SNR of 37.4 dB and completely suppressing the noise sources at 60Hz and at low frequencies; while the LMS algorithm exhibits an improvement in SNR of 33 dB and noisier spectrum at low frequencies. The GA algorithm is shown to be able to recover a heart signal with the QRS and T features being easily extracted. On the other hand, the LMS algorithm can also recover the input signal, however, with a lower SNR improvement and noisy QRS complex and T wave.
Language: English
Topic: ECG Monitoring and Analysis
Subfield: Cardiology and Cardiovascular Medicine
Field: Medicine
Domain: Health Sciences
Open Access status: green
OpenAlex ID: W264451053