Shock rates in an unselected real world cohort of patients with medtronic protecta implantable cardioverter defibrillators

Year: 2017
Type: article
Study design: Observational Study
Abstract: Objective: To investigate rates of inappropriate shocks in patients with a Medtronic Protecta implantable cardioverter defibrillator device (Medtronic Inc., Minneapolis, MN, USA).Methods: Consecutive patients implanted with Protecta defibrillators at a single centre between April 2010 and July 2012 were studied retrospectively.Results: Follow-up data was available for 599 patients(84% male; age 64 ± 14 years, range 16-92; 49% secondary prevention indication; 42% cardiac resynchronisation therapy, 24% single chamber, and 34% dual chamber defibrillator devices).Mean follow-up was 19 ± 7 months.57 patients (9.5%) received a shock: 8 patients(1.3%)had inappropriate shocks; 50 patients(8.3%)had appropriate shocks.Of the 10 inappropriate shocks, seven were for supraventricular tachycardia; one for T-wave oversensing; one due to lead displacement; and one occurred after an appropriate shock for ventricular tachycardia.Conclusions: Our reported inappropriate shock rate over a 20-month follow-up period is lower than previously reported.
Authors: Rao A, Barker D, Ng Kam Chuen MJ, Hughes S, Cullen D, Sharpe A, Wright DJ