Year: 2017
Type: conference-abstract
Abstract: ResultsThe differences in leaf positions compared with film and light field are beyond 0.1 mm and 1 mm (light field edge detection has a much bigger uncertainty).The acquisition and analysis for one strip-test take less than 4 min. ConclusionThe methodology employed analyzes a MLC strip-test in an Elekta LINAC in a fast and accurate way.
Source: Radiotherapy and Oncology
Authors: M. Islam, M. Farrokhkish, Y. Wang, B. Norrlinger, R. Heaton, D. Jaffray
Institutions: University of Toronto, Princess Margaret Cancer Centre
Language: English
Subfield: Radiation
Field: Physics and Astronomy
Domain: Physical Sciences
Open Access status: hybrid
Estimated APC paid: $3,770
OpenAlex ID: W2735967813