Work

A cryogenic thermal source for detector array characterization

Year: 2017
Type: article
Abstract: We describe the design, fabrication, and validation of a cryogenically compatible quasioptical thermal source for characterization of detector arrays. The source is constructed using a graphite-loaded epoxy mixture that is molded into a tiled pyramidal structure. The mold is fabricated using a hardened steel template produced via a wire electron discharge machining process. The absorptive mixture is bonded to a copper backplate enabling thermalization of the entire structure and measurement of the source temperature. Measurements indicate that the reflectance of the source is <0.001 across a spectral band extending from 75 to 330 GHz.
Language: English
Open Access status: green
OpenAlex ID: W2738580655
FWCI: 1.049
Cites: 26
Cited by: 22
Related to: 10