Work

Fluorescence Assisted Portable Cell Counting System

Year: 2013
Type: article
Abstract: The availability of a portable cell culture system such as the OptiCell TM Nunc membranes (OPNM) could play a key role in the field of cell-based biosensing, especially in optical cell-based biosensing.Such a system shows the potential for integration of multiple measurement and imaging modules.The integration of one such measurement module, a fluorescence intensity monitoring system for cell counting, is presented here.The optical design and implementation involved with the portable fluorescenceassisted cell counting device (PFACCD) are evaluated and discussed.The role of different optical design parameters such as the incident optical power, emitted fluorescence signal, and geometrical parameters associated with the PFACCD are estimated based on its geometry.The implemented design of the PFACCD system was tested with samples containing different concentrations of labeled L929 fibroblasts, loaded into the OPNMs.A calibration curve involving labeled L929 cell concentration and fluorescence intensity from a fluorometer was used to evaluate the detected fluorescence from the PFACCD system.The PFACCD was not able to detect fluorescence from point-source samples, such as cells.However, fluorescence was detected with higher concentrations of soluble dye.The results obtained using different geometry and excitation modes are discussed along with future design improvements.iii
Institution: University of Akron
Language: English
Subfield: Biophysics
Open Access status: green
OpenAlex ID: W2739194122
FWCI:
Cites: 48
Cited by:
Related to: 10