Multiple-Fiber Collection System for Scanning Tunneling Microscope Light Emission Spectroscopy
Year: 2001
Type: article
Abstract: We have constructed an optical fiber system for collecting light emission (LE) from the gap between a sample and the tip of a scanning tunneling microscope (STM). The collection system consists of four optical fibers with a core diameter of 600 µm and a numerical aperture (NA) of 0.2. The four optical fibers point radially at the tip-sample gap from four directions. To demonstrate the performance of the collection system, we evaluated its effective solid angle of collection by measuring the STM-LE from a Au evaporated film. The effective solid angle was estimated to be 0.33±0.03 sr, which corresponds to about 2.6 times that of a single optical fiber determined by the NA (0.13 sr). Although the solid angle of collection did not increase by a factor of four, the increase of the collection solid angle will improve the detection limit by reducing the accumulation time for spectral measurements.
DOI: 10.1143/jjap.40.5450
Authors: Ryuichi Arafune, Kenji Sakamoto, Kazuyuki Meguro, Minoru Satoh, Atsushi Arai, Sukekatsu Ushioda
Language: English
Subfield: Atomic and Molecular Physics, and Optics
Field: Physics and Astronomy
Domain: Physical Sciences
Open Access status: bronze
OpenAlex ID: W2064634170