Feasibility of 7T MRI for imaging fascicular structures of peripheral nerves

Year: 2017
Type: article
Abstract: INTRODUCTION: Evaluation of the nerve fascicular structure can be useful in diagnosing nerve damage, but it is a very challenging task with 3T MRI because of limited resolution. In this pilot study, we present the feasibility of high-resolution 7T MRI for examining the nerve fascicular structure. METHODS: A 3-dimensional (3D) gradient-spoiled sequence was used for imaging peripheral nerves in extremities. Images acquired with different in-plane resolutions (0.42 × 0.42 mm vs. 0.12 × 0.12 mm), and different main field strengths (7T vs. 3T) were compared. RESULTS: The individual nerve fascicles were identified at 0.12 × 0.12 mm resolution in both field strengths but not at 0.42 × 0.42 mm resolution. The fascicular structure was more sharply depicted in 7T images than in 3T images. DISCUSSION: High-resolution 3D imaging with 7T MRI demonstrated feasibility for imaging nerve fascicular structures. Muscle Nerve 57: 494-498, 2018.