Pneumatic low-coherence interferometry otoscope to quantify tympanic membrane mobility and middle ear pressure
Year: 2018
Type: article
Abstract: test. As a result, the compliance represents pneumatic TM mobility, and the amplitude ratio estimates MEP. The presence of a middle ear effusion (MEE) significantly decreased compliance (p<0.001). The amplitude ratio of the OM group was statistically less than that of the normal group (p<0.01), indicating positive MEP. Unlike tympanometry, pneumatic LCI otoscopy quantifies TM mobility as well as MEP regardless of MEE presence. With combined benefits of pneumatic otoscopy and tympanometry, pneumatic LCI otoscopy may provide new quantitative metrics for understanding TM dynamics and diagnosing OM.
Source: Biomedical Optics Express
DOI: 10.1364/boe.9.000397
Language: English
Topic: Ear Surgery and Otitis Media
Subfield: Otorhinolaryngology
Field: Medicine
Domain: Health Sciences
Open Access status: gold
Estimated APC paid: $1,913
Awards: G6854890528, G9000020026