The effect of wrist cuff width on oscillometric blood pressure waveforms
Year: 2008
Type: article
Abstract: Accuracy of blood pressure (BP) measurement with any method is affected by cuff size. Narrow cuff\noverestimates BP and wide cuff underestimates BP. Wrist cuff oscillometric BP monitors typically use narrow\ncuffs with bladders about 6 cm wide. The segment of the radial artery compressed by the wrist cuff presents a\nhydraulic resistance (Rc) to blood flow during BP measurement. Narrow cuff presents smaller Rc than wide cuff.\nChanges in Rc alter the amplitude envelope of oscillometric waveforms. Narrow cuff allows blood flow under\nthe cuff and into the hand sooner than wide cuff. The objective of this study was to demonstrate that narrow cuff\nalters the amplitude envelope in a manner that results in increased BP values when characteristic ratio algorithm\nis used. A specialized experimental system was used to acquire wrist oscillometric BP data from a narrow (6 cm)\ncuff and from a wide (10 cm) cuff. Data were acquired from 10 volunteers in sitting position. The results\nconfirmed the objective of the study. All narrow cuff BPs were higher than either the reference BPs or wide cuff\nBPs. Wide cuff algorithmic BPs were similar to the reference BPs. Wider wrist cuffs appear to be desirable for\nimproved accuracy and reliability of wrist BP monitors.
Authors: J. O. Jílek, Milan Štork
Language: English
Subfield: Biomedical Engineering
Field: Engineering
Domain: Physical Sciences
Open Access status: green
OpenAlex ID: W2788986676