Work

Myocardial infarction caused by active-fixation atrial pacing lead screwed into the anterior descending coronary artery

Year: 2018
Type: article
Study design: Case Report
Abstract: Coronary artery injury is a particularly rare complication of pacemaker implantation. Herein, we reported the case of a 76-year-old woman with sick sinus syndrome undergoing permanent pacemaker implantation who presented with chest pain a day after the procedure. During angiography, a pacemaker lead screwed into the left anterior descending coronary artery was found. The coronary artery was covered by a stent and the lead position was corrected. Based on the published research, the majority of right ventricular pacemaker leads implanted on the septum are indeed implanted on the anteroseptal junction. Since, the left anterior descending coronary artery covers this site, the risk of artery injury is higher.
Institution: Vilnius University
Language: English
Field: Medicine
Open Access status: diamond
OpenAlex ID: W2937941997
FWCI: 0.172
Cites:
Cited by: 3
Related to: 9