Work

iCAST Balloon-Expandable Covered Stent for Iliac Artery Lesions: 3-Year Results from the iCARUS Multicenter Study

Year: 2019
Type: article
Study design: Clinical Trial
Abstract: PURPOSE: To evaluate safety and effectiveness of the iCAST Covered Stent for treatment of iliac artery atherosclerotic lesions. MATERIALS AND METHODS: The iCARUS trial (ClinicalTrials.gov Identifier: NCT00593385) was a single-arm, prospective, multicenter study that enrolled 152 per protocol subjects at 25 sites in the United States and Germany. Subjects with multiple lesions and/or stents were eligible. The primary endpoint was the composite rate of death within 30 days, target lesion revascularization (TLR) within 9 months, or restenosis at 9 months after procedure. Secondary endpoints included major adverse vascular events (MAVEs), primary patency, freedom from TLR, and clinical success. RESULTS: Device and acute procedural success were achieved in 98.7% and 92.7% of cases, respectively. MAVE rate was 4.6% at 30 days. The 9-month primary composite endpoint rate was 8.1% (10/123), which was below the performance goal of 16.57%. Nine-month primary patency, defined as continuous flow without revascularization, bypass, or target limb amputation, was 96.4%. Freedom from TLR at 9 months and 3 years was 97.2% and 86.6%, respectively. Early clinical success was seen in 88.7% of subjects at 30 days with sustained clinical benefit in 72.4% of subjects at 3 years. CONCLUSIONS: The iCARUS study demonstrated that the iCAST Covered Stent was safe and effective for treatment of atherosclerotic iliac artery lesions with sustained clinical benefit out to 3 years.
Language: English
Subfield: Surgery
Field: Medicine
Open Access status: hybrid
Estimated APC paid: $3,000
OpenAlex ID: W2942066717
FWCI: 2.811
Cites: 30
Cited by: 25
Related to: 10