Quadrature Rotating-Frame Gradient Fields for Ultra-Low Field Nuclear Magnetic Resonance and Imaging

Year: 2005
Type: article
Abstract: Magnetic resonance imaging (MRI) in very low fields isfundamentally limited by untruncated concomitant gradients which causesevere distortions in image acquisition and volume selection if thegradient fields are strong compared to the static field. In this paper,it is shown that gradient fields oscillating in quadrature can be usedfor spatial encoding in low fields and provide substantial improvementsover conventional encoding methods using static gradients. In particular,cases where the gradient field is comparable to or higher than theexternal field, Gmax/B0>1, are examined. It is shown thatundistorted slice selection and image encoding is possible because ofsmaller geometric phase errors introduced during cyclic motions of theHamiltonian. In the low field limit (Gmax/B_0 ->infinity) sliceselection is achieved with a combination of soft pulse segments and acoherent train of hard pulses to average out concomitant fields over thefast scale of the rf Hamiltonian.