Application study of self-made standard dust in atmospheric particulate monitor calibration device

Year: 2019
Type: conference-paper
Abstract: The ISO12103-1 A2 test dust was used as the reference in the calibration device of the atmospheric particle monitor to investigate the performance of the self-made standard dust in this study. The result indicated that when the aerosol concentration of the calibration device was 500μg/m3,400μg/m3, 200μg/m3,100μg/m3 and 50μg/m3, the stability of self-made standard dust was close to the ISO12103-1 A2 test dust. In addition, when the aerosol with high concentration of 500μg/m3, medium concentration of 200μg/m3 and low concentration of 50μg/m3, the homogeneity and obtained concentration of the self-made standard dust and the ISO12103-1 A2 test dust evaluated by filter membrane sampling were not much differencd. Therefore, the self-made standard dust can replace ISO12103-1 A2 test dust to be used in the calibration device of particle monitor.