75-degree non-mydriatic single-volume optical coherence tomographic angiography
Year: 2019
Type: article
Abstract: Optical coherence tomography (OCT) and OCT angiography (OCTA) enable three-dimensional, high-resolution imaging of the eye. Yet, while they provide unprecedented structural and angiographic imaging detail, both have only limited fields of view in comparison to other imaging modalities like fundus photography. In this paper, we present a high-speed, high-sensitivity, swept source laser-based system that can acquire non-mydriatic 75-degree field of view OCT and OCTA images in a single complete scan without resorting to montaging techniques. The system uses an optimized scanning protocol and achieves capillary-level image quality. Such data may improve early detection of pathology and provide valuable information during disease monitoring.
Source: Biomedical Optics Express
Authors: Xiang Wei, Tristan T. Hormel, Yukun Guo, Yali Jia
Institution: Oregon Health & Science University
Language: English
Subfield: Biomedical Engineering
Field: Engineering
Domain: Physical Sciences
Open Access status: gold
Estimated APC paid: $1,913