Characterization of an Electronic Corneal Prosthesis System
Year: 2019
Type: article
Abstract: PURPOSE: Corneal opacity is a leading cause of reversible blindness worldwide. An electronic corneal prosthesis, or intraocular projector, could potentially restore high-quality vision without need for corneal clarity. MATERIALS AND METHODS: Four intraocular projection systems were constructed from commercially available electronic components and encased in biocompatible plastic housing. They were tested for optical properties, biocompatibility, heat dissipation, waterproofing, and accelerated wear. A surgical implantation technique was developed. RESULTS: . Temperature testing demonstrated less than 2°C increase in temperature after 1 h. Three devices lasted over 12 weeks under accelerated wear conditions. Implantation surgery was demonstrated via corneal trephination insertion in a cadaver eye. CONCLUSION: This is the first study to demonstrate and characterize fully functional intraocular projection systems. This technology has the potential to be an important new tool in the treatment of intractable corneal blindness.
Source: Current Eye Research
Language: English
Subfield: Radiology, Nuclear Medicine and Imaging
Field: Medicine
Domain: Health Sciences
Open Access status: green
OpenAlex ID: W2997465216