Year: 1981
Type: report
Abstract: Five-hole pitot tubes are widely used to determine directions and magnitudes of velocities in three-dimensional flow fields, because of their simplicity in handling and their reliability. This paper describes a method of reducing data obtained from five-hole pitot tube measurments with the aid of a few sets of calibration data. By using mini-computers, pitch and yaw angles and Mach numbers of flows can be computed simultaneously by this method with reasonable accuracy in the range of the pitot tube calibration.
Author: Hiroki Nishimura
Language: English
Subfield: Instrumentation
Field: Physics and Astronomy
Domain: Physical Sciences
Open Access status: green