Comparison of Abbott ID Now and Abbott m2000 Methods for the Detection of SARS-CoV-2 from Nasopharyngeal and Nasal Swabs from Symptomatic Patients
Year: 2020
Type: article
Abstract: COVID-19, NAAT, diagnostic testing, point of care T he ID Now COVID-19 (IDNCOV) assay performed on the ID Now instrument (AbbottDiagnostics, Inc., Scarborough, ME) is a rapid diagnostic test that can be performed in a point-of-care setting equivalent to Clinical Laboratory Improvement Amendments (CLIA)-waived testing.The assay utilizes isothermal amplification and can reportedly deliver results in approximately 5 to 13 min.As this assay could provide significant improvements to workflow in our hospital system, we sought to compare the performance of this test with our current coronavirus disease 2019 (COVID-19) assay, the Abbott RealTime SARS-CoV-2 (severe acute respiratory syndrome coronavirus 2) (ACOV) assay performed on the Abbott m2000 system (Abbott Molecular Inc., Des Plaines, IL).We compared the results from 524 paired foam nasal swabs (NS) tested on IDNCOV with nasopharyngeal swabs (NPS) placed in viral transport media tested on ACOV collected consecutively from symptomatic patients meeting current criteria for a diagnosis of COVID-19 (1).Five locations were included in the evaluation including three emergency departments (ED) and two immediate care centers (IMCC).IMCC A and ED 2 were experienced users of the IDNow platform.The other sites were new users of the platform and received training specifically for the IDNCOV.All ACOV testing was performed by one central clinical laboratory, and all NPS were heat inactivated for 30 min at 60°C prior to testing.NS were tested directly on the IDNCOV from IMCCs, and the tests were performed on-site.NS from the EDs were transported to the clinical microbiology laboratory in sterile transport containers (urine cups or conical tubes) and tested by laboratory personnel at each separate location.Statistical analysis was performed using SPSS v.26.The overall positivity rate in this sample collection was 35%, ranging from 22% to 60% among the five sites.Overall agreement was 75% positive agreement (95% confidence interval [95% CI], 67.74%, 80.67%) and 99% negative agreement (95% CI, 97.64%, 99.89%) between IDNCOV and ACOV for all specimens tested.Agreement at individual sites varied (Table 1).Two subjects tested positive on IDNCOV that were initially negative on ACOV.In case one, a repeat sample was positive on ACOV (repeat IDNCOV was not performed), and the case was resolved as a true positive result.For case two, all repeat testing (both IDNCOV and ACOV) was negative and was resolved as a likely false-positive result.This sample was collected during the first day of testing and could have been operator error.Fleiss kappa analysis comparing the performance at each of the sites demonstrated that strength of agreement between the sites (Table 1) was rated as good to very good with comparable standard errors.We interpret this to mean that a site's ability to run Citation
Source: Journal of Clinical Microbiology
DOI: 10.1128/jcm.00798-20
Language: English
Subfield: Infectious Diseases
Field: Medicine
Domain: Health Sciences
Open Access status: hybrid
OpenAlex ID: W3020248454