Year: 2010
Type: conference-paper
Abstract: CIMS Balloon Pump : A Novel Left Ventricle Assist Device (LVAD)5) Method A prototype balloon pump was developed -the outer body has a hard casing with a flexible silicone membrane inside it .The CIMS balloon pump prototype is placed just distal of the aortic valve in a human mock circulatory loop (MCL). The MCL is capable of simulating a variety of heart condition by adjusting left ventricle (LV) silicone sack contraction and peripheral resistance. Other than systemic circulation simulating LV and aortic flow, it is also equipped with left coronary artery (LCA) circulation. The LCA tube is placed just distal of the aortic valve and consists of a flow sensor, compliance syringe, pressure transducer and systolic resistor to simulate systolic compression of myocardium affecting the LCA.The CIMS balloon pump is connected to an IABP driver console. Once HF condition is set on the MCL the CIMS balloon pump is activated at 1:1 ratio i.e. at every heart beat triggered at the dicrotic notch. Data was taken for non-assisted flow and assisted flow for 40 seconds. Several parameters were measured and student t-test was used to determine the significance of flow augmentation.
Language: English
Subfield: Cardiology and Cardiovascular Medicine
Field: Medicine
Domain: Health Sciences
Open Access status: green
Funder: Deutsche Forschungsgemeinschaft
Awards: G2113323751, G3513178703, G7764990065
OpenAlex ID: W30253557
