I. An isothermal titration microcalorimeter. II. Importance of micronutrients in nutrition. III. Zinc bibliography
Year: 1973
Type: article
Abstract: I. An isothermal titration microcalorimeter having a colume of 4 ml and capable of temerature control to ±2 x 10-5°C is described. Major components include a constant temperature water bath controlled to ±3 x 10-4°C, a platinum reation vessel, and an isothermal control circuit consisting of constant Peltier thermoelectric cooling and variable Joule heating controlled by a thermistor in an AC Wheatstone bridge circuit. The calorimeter was tested by measuring the heat of ionization of water and was found to produce data accurate to ± 0.1% where small samples are used such as in the investigation of many biological systems.
Author: John Willard Gardner
Language: English
Subfield: Physical and Theoretical Chemistry
Field: Chemistry
Domain: Physical Sciences
Keywords: micronutrients, nutrition, zinc, Joule heating