A Comparison Of Manual Hemocytometry And Different Modes Of Mindray BC-6800 Hematology Analyzer For Cell Count In Peritoneal And Pleural Fluids
Year: 2020
Type: article
Abstract: Objective: Total white blood cell (WBC) count in body fluids (BF) is important in diagnosis and treatment of inflammatory and infectious diseases.We aimed to compare hemogram and BF modes of Mindray BC-6800 hematology analyzer with the manual hemocytometry method for determination of total WBC count in peritoneal and pleural fluids.Method: Our study consisted of a total of 143 specimens, 109 peritoneal fluid and 34 pleural effusion fluid.Each sample was analyzed consecutively twice, first with the manual method and, later with both hemogram mode and the BF mode of Mindray BC-6800 automated hematology analyzer.Results: There was a statistically significant difference between manual count and hemogram mode (for peritoneal fluids p< 0.001, for pleural fluids p< 0.001).There was a statistically significant difference between the hemogram mode and BF mode too.(for peritoneal fluids p< 0.001, for pleural fluids p< 0.001).However, there was no statistically significant difference between the manual count and BF mode (for peritoneal fluids p=0.236, for pleural fluids p=0.627).Less than100 cells/mL, there was a statistically significant difference between each of the counting methods (Manual count vs hemogram mode p< 0.001, manual count vs BF mode p= 0.012, hemogram mode vs BF mode p< 0.001).
Authors: Ünsal SAVCI, Mustafa ŞAHİN, Barış ESER
Language: English
Subfield: Physiology
Field: Medicine
Domain: Health Sciences
Open Access status: diamond
OpenAlex ID: W3036113889