Glucose as the Fifth Vital Sign: A Randomized Controlled Trial of Continuous Glucose Monitoring in a Non-ICU Hospital Setting
Year: 2020
Type: article
Study designs: Randomized Controlled Trial, Clinical Trial
Abstract: OBJECTIVE The current standard for hospital glucose management is point-of-care (POC) testing. We conducted a randomized controlled trial of real-time continuous glucose monitoring (RT-CGM) compared with POC in a non–intensive care unit (ICU) hospital setting. RESEARCH DESIGN AND METHODS A total of 110 adults with type 2 diabetes on a non-ICU floor received RT-CGM with Dexcom G6 versus usual care (UC). RT-CGM data were wirelessly transmitted from the bedside. Hospital telemetry monitored RT-CGM data and notified bedside nursing of glucose alerts and trends. Standardized protocols were used for interventions. RESULTS The RT-CGM group demonstrated significantly lower mean glucose (M∆ = −18.5 mg/dL) and percentage of time in hyperglycemia >250 mg/dL (−11.41%) and higher time in range 70–250 mg/dL (+11.26%) compared with UC (P values <0.05). Percentage of time in hypoglycemia was very low. CONCLUSIONS RT-CGM can be used successfully in community-based hospital non-ICU settings to improve glucose management. Continuously streaming glucose readings may truly be the fifth vital sign.
Source: Diabetes Care
DOI: 10.2337/dc20-1016
Institution: Scripps Whittier Diabetes Institute
Language: English
Subfield: Endocrinology, Diabetes and Metabolism
Field: Medicine
Domain: Health Sciences
Open Access status: gold
OpenAlex ID: W3082942486