Impact of Patient Alignment on Image Quality in C-Arm Computed Tomography – Evaluation Using an ACR Phantom
Year: 2020
Type: article
Abstract: PURPOSE: To evaluate the influence of patient alignment and thereby heel effect on the image quality (IQ) of C-arm flat-panel detector computed tomography (CACT). MATERIALS AND METHODS: An ACR phantom placed in opposite directions along the z-axis (setup A and B) on the patient support was imaged using CACT. Image acquisition was performed with three different image acquisition protocols. The images were reconstructed with four convolution kernels. IQ was assessed in terms of high contrast using the modulation transfer function (MTF) and low contrast by assessing the image noise, signal-to-noise ratio (SNR) and contrast-to-noise ratios (CNR) as well as the reliability of density measurements. Furthermore, the dose intensity profiles were measured free-in-air. RESULTS: The MTF in setup B is higher than the MTF measured in setup A (p < 0.01). The image noises measured in setup A for the air and bone inserts were higher compared to those measured in setup B (p > 0.05). Opposite behavior has been observed for the polyethylene, water-equivalent and acrylic inserts. The SNR for all inserts is inversely related to the image noise. A systematically increasing or decreasing trend of CNR could not be observed (p > 0.05). The intensity profile measured by the detector system free-in-air showed that the anode heel effect is perpendicular to the z-axis. CONCLUSION: The patient alignment has a minor influence on the IQ of CACT. This effect is not based on the X-ray anode heel effect but is caused mainly by the non-symmetrical rotation of CACT. KEY POINTS: · The impact of patient alignment and thereby the heel effect on the image quality of CACT was analyzed.. · The patient alignment has a minor influence on the physical parameters related to image quality, such as noise, SNR, and MTF.. · This effect is based mainly on the non-symmetrical rotation of CACT.. CITATION FORMAT: · Alikhani B, Renne J, Maschke S et al. Impact of Patient Alignment on Image Quality in C-Arm Computed Tomography - Evaluation Using an ACR Phantom. Fortschr Röntgenstr 2021; 193: 417 - 426.
DOI: 10.1055/a-1238-2802
Authors: Babak Alikhani, Julius Renne, Sabine Maschke, Jan B. Hinrichs, Frank K. Wacker, Thomas Werncke
Institutions: Diakovere, Medizinische Hochschule Hannover
Language: English
Topic: Radiation Dose and Imaging
Subfield: Radiology, Nuclear Medicine and Imaging
Field: Medicine
Domain: Health Sciences
Open Access status: bronze
OpenAlex ID: W3083090504