Work

Further development of IDGS: Isotope dilution gamma-ray spectrometry

Year: 1991
Type: article
Abstract: The isotope dilution gamma-ray spectrometry (IDGS) technique for determining the plutonium concentration and isotopic composition of highly radioactive spent-fuel dissolver solutions has been further developed. Both the sample preparation and the analysis have been improved. The plutonium isotopic analysis is based on high-resolution, low-energy gamma-ray spectrometry. The plutonium concentration in the dissolver solutions then is calculated from the measured isotopic differences among the spike, the dissolver solution, and the spiked dissolver solution. Plutonium concentrations and isotopic compositions of dissolver solutions analyzed from this study agree well with those obtained by traditional isotope dilution mass spectrometry (IDMS) and are consistent with the first IDGS experimental result. With the current detector efficiency, sample size, and a 100-min count time, the estimated precision is {approximately}0.5% for {sup 239}Pu and {sup 240}Pu isotopic analyses and {approximately}1% for the plutonium concentration analysis. 5 refs., 2 figs., 7 tabs.
Language: English
Subfield: Radiation
Open Access status: green
OpenAlex ID: W3087920
FWCI:
Cites: 1
Cited by: 3