High-Energy Calibration of a BGO Detector of the GLAST Burst Monitor

Year: 2011
Type: article
Abstract: The understanding of the instrumental response of the GLAST Burst Monitor BGO detectors at energies above the energy range which is accessible by common laboratory radiation sources (< 4.43 MeV), is important, especially for the later cross-calibration with the LAT response in the overlap region between {approx}20 MeV to 30 MeV. In November 2006 the high-energy calibration of the GBM-BGO spare detector was performed at the small Van-de-Graaff accelerator at SLAC. High-energy gamma-rays from excited {sup 8}Be* (14.6 MeV and 17.5 MeV) and {sup 16}O* (6.1 MeV) were generated through (p, {gamma})-reactions by irradiating a LiF-target. For the calibration at lower energies radioactive sources were used. The results, including spectra, the energy/channel-relation and the dependence of energy resolution are presented.