Year: 2020
Type: article
Abstract: Progress in electromagnetic induction imaging with atomic magnetometers has brought its domain to the edge of the regime useful for biomedical imaging. However, a demonstration of imaging below the required 1 Sm−1 level is still missing. In this Letter, we use an 87Rb radio frequency atomic magnetometer operating near room temperature in an unshielded environment to image calibrated solutions mimicking the electric conductivity of live tissues. By combining the recently introduced near-resonant imaging technique with a dual radio frequency coil excitation scheme, we image 5 ml of solutions down to 0.9 Sm−1. We measure a signal-to-noise ratio of 2.7 at 2 MHz for 0.9 Sm−1, increased up to 7.2 with offline averaging. Our work is an improvement of 50 times on previous imaging results and demonstrates the sensitivity and stability in unshielded environments required for imaging biological tissues, in particular for the human heart.
Source: Applied Physics Letters
DOI: 10.1063/5.0002146
Authors: Cameron Deans, Luca Marmugi, Ferruccio Renzoni
Institution: University College London
Language: English
Subfield: Atomic and Molecular Physics, and Optics
Field: Physics and Astronomy
Domain: Physical Sciences
Open Access status: green
OpenAlex ID: W3014958530