Portable Oxygen-Sensing Device for the Improved Assessment of Compartment Syndrome and other Hypoxia-Related Conditions
Year: 2020
Type: article
Abstract: Measurement of intramuscular oxygen could play a key role in the early diagnosis of acute compartment syndrome, a common condition occurring after severe trauma leading to ischemia and long-term consequences including rhabdomyolysis, limb loss, and death. However, to date, there is no existing oxygen sensor approved for such a purpose. To address the need to improve the assessment of compartment syndrome, a portable fiber-optic device for intramuscular oxygen measurements was developed. The device is based on phosphorescence quenching, where the tip of an optical fiber was coated with a poly(propyl methacrylate) (PPMA) matrix containing a brightly emitting Pt(II)-core porphyrin. The optoelectronic circuit is highly portable and is based on a microspectrometer and a microcontroller readout with a smartphone. Results from an in vivo tourniquet porcine model show that the sensor is sensitive across the physiological oxygen partial pressure range of 0–80 mmHg and exhibits an appropriate and reproducible response to changes in intramuscular oxygen. A commercial laboratory oxygen sensor based on a lifetime measurement did not respond as expected.
Source: ACS Sensors
Institutions: Harvard University, Massachusetts General Hospital
Language: English
Subfield: Bioengineering
Field: Chemical Engineering
Domain: Physical Sciences
Open Access status: green
Award: G3199007678
OpenAlex ID: W3112599550