Year: 2020
Type: article
Abstract: Indonesian Nuclear Energy Regulatory Agency (Bapeten) has set the head CT DRL value of 50 mGy. CT parameter settings will affect the dose and image quality. Tube current, rotation time, and pitch parameters that used in head CT protocol with infarction case in GoetengTaroenadibrata General Hospital produced a displayed CTDIvol of 78, 7 mGy. The aim of this study was to obtain a head CT protocol with infarction cases that are able to produce optimal image quality and radiation doses that do not exceed DRL. This study was conducted by applying four head CT protocols with tube current, rotation time, and pitch variation with the monitor’s CTDIvol value does not exceed DRL in the low contrast resolution and spatial resolution phantom, also in four sample groups, each of which numbered 15 patients with clinical infarction in which each group received only one treatment. Measurement of physical image quality is done by measuring noise and SNR using the ROI software, and low contrast resolution and spatial resolution which were assessed visually by three medical physicists. Clinical quality was carried out by assessing anatomic and pathological information by three radiologists. Head CT protocol with parameters of 250 mA, rotation time 0.75 seconds, and pitch 0,750 produces the best physical and clinical image quality also radiation reduction up to 53, 8 %.