Thermal experiment of silicon PIN detector

Year: 2008
Type: article
Abstract: The experiment of this paper is the thermal test of the leakage current of silicon PIN detector. Raising temperature may cause the detector to increase leakage current, decrease depletion and increase noise. Three samples are used in the experiment. One (called ΔE) is the sample of 100 μm in thickness. The other two (called E 1 and E 2 ) are stacks of five detectors of 1000 μm in thickness. All of them are 12 mm in diameter. The experiment has been done for 21 hours and with power on continuously. The samples have undergone more than 60°C for about one hour. They are not degenerated when back to the room temperature. The depletion rate is temperature and bias voltage related. With the circuit of the experiment and temperature at 35°C, ΔE is still depleted while E 1 and E 2 are 94.9% and 99.7% depleted respectively. The noises of the samples can be derived from the values at room temperature and the thermal dependence of the leakage currents. With the addition of the noise of the pre-amplifier, the noises of E 1 , E 2 and ΔE at 24°C are 16.4, 16.3, and 10.5 keV (FWHM) respectively while at 35°C are about 33.6, 33.1, and 20.6 keV (FWHM) respectively.