Year: 2012
Type: article
Abstract: The gas gain and energy resolution of single and double THGEM detectors (5 cm × 5 cm effective area) with mini-rims (rim less than 10 μ m) were studied. The maximum gain was found to reach 5 × 10 3 and 2 × 10 5 for single and double THGEMs respectively, while the energy resolution for 5.9 keV X-rays varied from 18% to 28% for both single and double THGEM detectors of different hole sizes and thicknesses. Different combinations were also investigated of noble gases (argon, neon) mixed with a quantity of other gases (isobutane, methane) at atmospheric pressure.
Source: Chinese Physics C
Language: English
Subfield: Nuclear and High Energy Physics
Field: Physics and Astronomy
Domain: Physical Sciences
Open Access status: bronze
OpenAlex ID: W1606717166