Aflojamiento de los cotilos roscados en las prótesis de cadera
Year: 2000
Type: article
Study design: Observational Study
Abstract: We report our long-term results using a threaded acetabular cup (Lord, Howmedica) in 145 hips with a follow-up longre than 10 years. Ten patients died during this period of time (12 cups) and we have reviewed 65 implants with a mean age at operation of 60 years. The clinical results according to Merle d'Aubigne score, showed that the hip is still pain free, but there was a loss in walking ability of 0,6 pints (p<0.001). fourteen cups were revised for aseptic loosening (21%) and revision was pending at the most recent follow-up examination in additional two cups. The overall percentage of symptomatic migration was 24%. We conclude that long term stability of threaded cups is not warranted and different factors can influence the revision rate (design, biomechanic and technical factors), indicating that further use of these acetabular components cannot be recommended.
Author: Javier Vaquero
Language: Spanish
Subfield: Surgery
Field: Medicine
Domain: Health Sciences
Open Access status: green
OpenAlex ID: W31568853