Magnetic resonance angiography and perfusion mapping by arterial spin labeling using Fourier transform–based velocity‐selective pulse trains: Examination on a commercial perfusion phantom
Year: 2021
Type: article
Abstract: PURPOSE: Benchmarking of flow and perfusion MR techniques on standardized phantoms can facilitate the use of advanced angiography and perfusion-mapping techniques across multiple sites, field strength, and vendors. Here, MRA and perfusion mapping by arterial spin labeling (ASL) using Fourier transform (FT)-based velocity-selective saturation and inversion pulse trains were evaluated on a commercial perfusion phantom. METHODS: The FT velocity-selective saturation-based MRA and FT velocity-selective inversion-based ASL perfusion imaging were compared with time-of-flight and pseudo-continuous ASL at 3 T on the perfusion phantom at two controlled flow rates, 175 mL/min and 350 mL/min. Velocity-selective MRA (VSMRA) and velocity-selective ASL (VSASL) were each performed with three velocity-encoding directions: foot-head, left-right, and oblique 45°. The contrast-to-noise ratio for MRA scans and perfusion-weighted signal, as well as labeling efficiency for ASL methods, were quantified. RESULTS: On this phantom with feeding tubes having only vertical and transverse flow directions, VSMRA and VSASL exhibited the dependence of velocity-encoding directions. The foot-head-encoded VSMRA and VSASL generated similar signal contrasts as time of flight and pseudo-continuous ASL for the two flow rates, respectively. The oblique 45°-encoded VSMRA yielded more uniform contrast-to-noise ratio across slices than foot-head and left-right-encoded VSMRA scans. The oblique 45°-encoded VSASL elevated labeling efficiency from 0.22-0.68 to 0.82-0.90 through more uniform labeling of the entire feeding tubes. CONCLUSION: Both FT velocity-selective saturation-based VSMRA and FT velocity-selective inversion-based VSASL were characterized on a commercial perfusion phantom. Careful selection of velocity-encoding directions along the major vessels is recommended for their applications in various organs.
Source: Magnetic Resonance in Medicine
DOI: 10.1002/mrm.28805
Institutions: Kennedy Krieger Institute, Johns Hopkins University
Language: English
Subfield: Radiology, Nuclear Medicine and Imaging
Field: Medicine
Domain: Health Sciences
Open Access status: hybrid
Estimated APC paid: $3,500
Awards: Development of MRI microvascular biomarkers in cognitive impairment and dementia, MRI-based Quantitative Brain Oximetry for Sickle Cell Disease, State of the Art 3T Research Scanner, Resource for Quantitative Functional FMRI, Non-Contrast-Enhanced Velocity-Selective MR Angiography at 3T for Cerebrovascular Diseases, Velocity-Selective Arterial Spin Labeling based Perfusion Mapping for Cerebrovascular Diseases, NON-CONTRAST-ENHANCED PERIPHERAL MR ANGIOGRAPHY, MRI Resource for Physiologic, Metabolic and Anatomic Biomarkers, (CIRCADIAN) Circadian Disruption as Mediator of Cardiometabolic Risk in Air Pollution
OpenAlex ID: W3159114765