Response to Comment on Lee et al. Multicenter, Head-to-Head, Real-World Validation Study of Seven Automated Artificial Intelligence Diabetic Retinopathy Screening Systems. Diabetes Care 2021;44:1168–1175
Year: 2020
Type: article
Abstract: We thank Dr. Soliz for his expedient comments (1) regarding our work evaluating seven artificial intelligence (AI) algorithms in a head-to-head fashion (2). We welcome the opportunity to clarify these points of concern.
We agree that the two U.S. Food and Drug Administration–approved diabetic retinopathy (DR) systems, one of which was tested in our study, use a threshold of moderate nonproliferative DR (NPDR) or higher for referral. We set the threshold to any DR versus no DR because that represents the clinical practice at the Veterans Affairs (VA) teleretinal system, as stated in our methods. All participating companies were aware of the study threshold prior to submitting their algorithms and were allowed to adjust their algorithms accordingly. As the VA manages one of the largest teleretinal screening systems in the U.S., …