蛍光x線分析法による岩石試料の定量化学分析(1)主成分元素
Year: 2017
Type: article
Abstract: The X-ray fluorescence (XRF) analyzer (Shimadzu XRF-1800) of the Division for Chronological Research, Institute for Space‒Earth Environmental Research, Nagoya University, was moved from the Furukawa Hall to the Research Institutes Building II on April 2016. To perform accurate and stable measurements of major elements with the XRF analyzer again after the moving, we examined the optimal analytical condition and remade calibration curves using 44 geochemical reference rock samples from the Geological Survey of Japan. The measurement result using the new calibration curves showed that the measured values were close to the reference values for all major elements except for ones with very low concentrations, indicating that accurate measurement of major elements has been enabled.
Source: 名古屋大学年代測定研究
DOI: 10.18999/nagubc.1.66
Language: Japanese
Subfield: Radiation
Field: Physics and Astronomy
Domain: Physical Sciences
Open Access status: green
OpenAlex ID: W3171998910