A practical guide to quartz crystal microbalance with dissipation monitoring of thin polymer films
Year: 2021
Type: article
Abstract: Quartz crystal microbalance with dissipation (QCM‐D) monitoring is a powerful tool used to sensitively examine the real‐time responses of polymer films to external responses. For example, the technique is commonly used to monitor film growth, material adsorption, thin film swelling, and ion exchange. With its rapidly expanding use, this review is intended to introduce new users to the basic principles of QCM‐D, along with practical challenges and remedies specific to polymer thin films. For both new and experienced users, specific case studies are highlighted including layer‐by‐layer assembly, electrochemical QCM‐D, swelling, sensing, and biological application. Last, the review recommends future directions for research and areas of growth.
Source: Journal of Polymer Science
DOI: 10.1002/pol.20210324
Institution: Texas A&M University
Language: English
Subfield: Electrical and Electronic Engineering
Field: Engineering
Domain: Physical Sciences
Open Access status: hybrid
Estimated APC paid: $3,200