Work

Evaluation of noninvasive blood pressure in anesthetized horses: Bias, limits of agreement, and comparative detection of a predetermined mean blood pressure warranting treatment.

Year: 2021
Type: article
Abstract: = 0.08). Wide limits of agreement were observed. Percentage errors were smaller for SAP (39%) than for MAP and DAP (48% and 72%). The oscillometric monitor indicated there was a requirement for blood pressure treatment with a true positive rate of 82%, consequently, it failed 18% of the times. The false positive rate (unnecessary treatment) was 55%.
Language: English
Subfield: Small Animals
Field: Veterinary
Open Access status: green
OpenAlex ID: W3197497455
FWCI: 0.263
Cites: 17
Cited by: 4
Related to: 9