Abstract: This paper describes the design of a continuous wave blood flow bidirectional Doppler system based on an open, high-resolution architecture that is portable and low-cost. The system incorporates the advantages of expensive systems with dedicated hardware. The system is composed of a flow detector probe, a signal conditioning stage, a direction detection module, a signal processing stage and a GUI. The direction detection of the signal is achieved using a fast digital Phasing Filter. The Doppler signal is processed by a Short-Time Fourier Transform-based algorithm. This is commonly used as a reference. Nevertheless, the system allows us to incorporate alternative high-resolution spectral estimation methods that might offer more precise information to the specialist
Authors: F. García, E. Moreno, J. Solano, Marina Esther Ahumada Barragán, Alfredo Robles Álvarez de Sotomayor, Manuel Fuentes
Language: English
Subfield: Biomedical Engineering
Field: Engineering
Domain: Physical Sciences
Open Access status: diamond